Intra-assay Precision (Precision within an assay): CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess.
Inter-assay Precision (Precision between assays): CV%<10%
Three samples of known concentration were tested in twenty assays to assess.
線性度:
To assess the linearity of the assay, samples were spiked with high concentrations of rat KI-67 in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay.
Sample
Serum(n=4)
1:1
Average %
87
Range %
83-90
1:2
Average %
93
Range %
90-96
1:4
Average %
89
Range %
85-93
1:8
Average %
100
Range %
95-105
回收率:
The recovery of rat KI-67 spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section.
Sample Type
Average % Recovery
Range
Serum (n=5)
89
85-93
EDTA plasma (n=4)
92
88-96
標準曲線:
These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed.